Pulse Processor Time
operator-selected value for the time designated to record a response by the detector.
Discussion—A higher value (longer time) results in a more accurate determination of the detector amplifier pulse height (better spectral resolution). A lower value results in a higher count rate but with reduced spectral resolution.